DT-6605****高压绝缘电阻测试仪是款手持式仪表,主要用来测量交流/直流电压、电阻、短路蜂鸣测试和绝缘电阻测量。
IEC 61010-1 (CAT IV 600V污染等级2级)
IEC 61010-031 (手持探头的要求)
绝缘电阻测量 |
||||
额定电压 |
500V |
1000V |
2500V |
5000V |
测量量程 (自动测量) |
0.005~6.000MΩ |
0.005~6.000MΩ |
0.05~60.00MΩ |
0.05~60.00MΩ |
6.01~60.00MΩ |
6.01~60.00MΩ |
60.1~600.0MΩ |
60.1~600.0MΩ |
|
60.1~600.0MΩ |
60.1~600.0MΩ |
0.601~6.00GΩ |
0.601~6.00GΩ |
|
0.601~6.00GΩ |
0.601~6.00GΩ |
6.01~60.0GΩ |
6.01~60.0GΩ |
|
开路电压 |
DC 500V +20%,-0% |
DC 1000V +20%,-0% |
DC 2500V +20%,-0% |
DC 5000V +20%,-0% |
额定电流 |
1~1.2mA (负载0.5 MΩ ) |
1~1.2mA (负载1 MΩ ) |
1~1.2mA (负载2.5 MΩ) |
1~1.2mA (负载为5 MΩ时,推荐使用电源适配器) |
短路电流 |
大约1mA |
|||
误差 |
+2.5%re*g+15digits (at 0.005~600.0MΩ); +3%re*g+15digits (at 0.601~6.00GΩ); +4%re*g+15digits (at 6.01~60.0GΩ); |
|||
电压控制量程 |
5~6000VDC (分辨率:1V);*度:+1.5%re*g+5digits |
|||
注意:此仪表用于检测保存于受试设备中的电荷是否被释放。LCD显示的测量电压值为参考值,请注意,当仪表连接外部AC或DC电压时,LCD所示电压值大于30V,非正确的数值。 |